The depth within the body, small diameter, long length, and varying tissue surrounding the spinal cord impose specific considerations when d
… (voir plus)esigning RF coils. The optimal coil configuration for 7 T cervical spinal cord MRI is unknown and currently there are very few coil options. The purpose of this work was (1) to establish a quality control protocol for evaluating 7 T cervical spinal cord coils, and (2) to use that protocol to evaluate the performance of four different coil designs.
Three healthy volunteers and a custom anthropomorphic phantom (the traveling spines cohort) were scanned at seven 7 T imaging centers using a common protocol and each center's specific cervical spinal cord coil. Four different coil designs were tested (two in‐house, one Rapid Biomedical, and one MRI.TOOLS design).
The Rapid Biomedical coil was found to have the highest B1+ efficiency, whereas one of the in‐house designs (NeuroPoly Lab) had the highest SNR and the largest spinal cord coverage. The MRI.TOOLS coil had the most uniform B1+ profile along the cervical spinal cord; however, it was limited in its ability to provide the requested flip angles (especially for larger individuals). The latter was also the case for the second in‐house coil (MSSM).
The results of this study serve as a guide for the spinal cord MRI community in selecting the most suitable coil based on specific requirements and offer a standardized protocol for assessing future coils.